USP 645 – easy and accurate conductivity calibration with an insacal master meter
The relationship between Insacal and Irish Power and Process appears to be mutually beneficial, combining Insacal’s expertise in conductivity calibration technology with IPP’s 30+ years of experience serving Irish industry in process instrumentation and calibration products8
enhanced accuracy
The indicator/transmitter is calibrated using a high precision resistor prior to the loop-calibration. e.g. Master displays 100,0uS/cm Ambient temperature = 30C Cell constant determined by a 100uS/cm standard solution (Ucc = 0.36%) Coverage factor = 2 (level of confidence at approximately 9%%) Standard resistor acc = 0.05%. Distribution: square Calibration of standard resistor: 100ppm Confidence interval: 2. Temperature influence of standard resistor 15ppm Distribution: square Total uncertainty = 0.37%